User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: Cypher E525: Strain Indicator DATE: 19-Dec-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 217 CONFIGURATION: Datron 1281 STANDARD: Measurements Group 1550A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD Measurements Group 1550A 1.004 DISP Required Equipment 1.004 DISP Measurements Group 1550A Strain Calibrator 1.005 HEAD INITIAL CONDITIONS 1.006 DISP Connect the UUT as follows: 1.006 DISP E525 INPUT 5 & 7 ................. 1550A P+ (350 ohm) 1.006 DISP E525 INPUT 6 & 8 ................. 1550A P- (350 ohm) 1.006 DISP E525 INPUT 9 ..................... 1550A S+ (350 ohm) 1.006 DISP E525 INPUT 10 & 11 ............... 1550A S- (350 ohm) 1.007 DISP Plug the UUT into an appropriate power source. 1.008 DISP On the 1550A set the MICROSTRAIN PER STEP switches 1.008 DISP to 0,0,0. Set the POLARITY to + and depress the 350 ohm 1.008 DISP CIRCUIT switch. 1.009 DISP On the UUT front panel simultaneously press the 1.009 DISP LIMITS and "4" keys (ENTER)> 1.010 DISP The display should read P= 0. 1.011 DISP Use the UP/DOWN arrow keys to set P= 73. Press ENTER. 1.012 OPBR Does the display read Clr? 1.013 JMPT 1.018 1.014 DISP Again, simultaneously press the LIMITS and "4" keys. 1.015 DISP This time set P= 4. Press ENTER. 1.016 OPBR Does the display read Clr? 1.017 JMPF 14.004 1.018 DISP Press the DOWN arrow 1 time. 1.019 DISP The display should read Count. If it does not, 1.019 DISP continue pressing the arrow keys until it does. 1.020 DISP Press LIMITS and "4" (ENTER). 1.021 DISP Write down the value in the display. This is necessary 1.021 DISP to restore the operational status after the calibration. 1.022 DISP Using the arrow keys set this value to 1. Press ENTER. 1.023 DISP The display should read F.SCALE. If not use arrow 1.023 DISP keys as necessary. 1.024 DISP Press ENTER. Write down displayed value. 1.025 DISP Use arrow keys the set to 30000. Press ENTER. 1.026 DISP The dispaly should read d.Point. Press ENTER. 1.027 DISP Note position of decimal point. Use the arrow keys 1.027 DISP for a display of 30000 with no decimals. Press ENTER. 1.028 DISP The display should read Filt. Press ENTER. 1.029 DISP Note the current filter setting. Use the arrow keys 1.029 DISP to set the filter to 6. Press ENTER. 1.030 DISP The display should read Ecn. Press ENTER. 1.031 DISP Note the current excitation value. Use the arrow keys 1.031 DISP to set the excitation to 1. Press ENTER. 1.032 DISP The display should read Aut.CAL. Press the DOWN arrow 1.032 DISP 1 time. The dispaly should read CAL. Press ENTER. 1.033 DISP The dispaly should read Pr.CAL. Press ENTER. 1.034 DISP The dispaly will momentarily displat Cal.Pt. Then 1.034 DISP it will display the current calibration point. 1.035 DISP Note the current cal point. Use the arrow keys to 1.035 DISP set the cal point to 30000. Press ENTER. 1.036 DISP Ensure that the strain calibrator is set to 0,0,0. 1.037 DISP Press the "1" key. Followed by pressing the "2" key. 1.038 DISP Set the strain calibrator to 6,0,0. 1.039 DISP Press the "4" key. The display shoul read 30000. 1.039 DISP Press ENTER. Press UP arrow and "1" (RESET). 1.040 HEAD {* Linearity Tests *} 1.041 JMP 2.001 1.042 EVAL 2.001 MATH m[1] = 30000 2.002 MATH m[2] = 25000 2.003 MATH m[3] = 20000 2.004 MATH m[4] = 15000 2.005 MATH m[5] = 10000 2.006 MATH m[6] = 5000 2.007 MATH m[7] = 2500 2.008 MATH m[8] = 500 2.009 MATH m[9] = 250 2.010 MATH m[10] = 100 # 2.011 DISP Set the 1550A MICROSTRAIN buttons to 6,0,0. 2.011 MEMI Enter the UUT indication. 2.012 MEME >11 2.013 DISP Set the 1550A MICROSTRAIN buttons to 5,0,0. 2.014 MEMI Enter the UUT indication. 2.015 MEME >12 2.016 DISP Set the 1550A MICROSTRAIN buttons to 4,0,0. 2.017 MEMI Enter the UUT indication. 2.018 MEME >13 2.019 DISP Set the 1550A MICROSTRAIN buttons to 3,0,0. 2.020 MEMI Enter the UUT indication. 2.021 MEME >14 2.022 DISP Set the 1550A MICROSTRAIN buttons to 2,0,0. 2.023 MEMI Enter the UUT indication. 2.024 MEME >15 2.025 DISP Set the 1550A MICROSTRAIN buttons to 1,0,0. 2.026 MEMI Enter the UUT indication. 2.027 MEME >16 2.028 DISP Set the 1550A MICROSTRAIN buttons to 0,5,0. 2.029 MEMI Enter the UUT indication. 2.030 MEME >17 2.031 DISP Set the 1550A MICROSTRAIN buttons to 0,1,0. 2.032 MEMI Enter the UUT indication. 2.033 MEME >18 2.034 DISP Set the 1550A MICROSTRAIN buttons to 0,0,5. 2.035 MEMI Enter the UUT indication. 2.036 MEME >19 2.037 DISP Set the 1550A MICROSTRAIN buttons to 0,0,2. 2.038 MEMI Enter the UUT indication. 2.039 MEME >20 2.040 CALL Least Squares Sub 2.041 CALL Strain Linearity Sub 2.042 HEAD {} 2.043 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 2.044 HEAD {º Analoge Output Tests º} 2.045 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 2.046 HEAD ÍÍ Analoge Output Tests ÍÍ 2.047 JMP 3.001 2.048 EVAL 3.001 DISP On the rear of the UUT locate terminals 11, 12, and 13 3.001 DISP on the 01 terminal strip. 3.002 OPBR Is there a connection made at terminal 11? 3.003 JMPT 8.002 3.004 DISP Remove the connections at terminals 12 and 13. Be sure 3.004 DISP to note which wire comes from which terminal. 3.005 DISP ÿÿÿÿÿÿÿÿConnect the UUT as follows: 3.005 DISP terminal 13 ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]Datron I+ 3.005 DISP terminal 12 ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]Datron I- 3.006 DISP Set the 1550A MICROSTRAIN buttons to 5,0,0. 3.007 IEEE [@1281]DCI AUTO 3.008 ACC 200 20.000mA 100P% 20P/ 3.009 IEEE [@1281]RDG?[I] 3.010 MEME 3.011 MEMC mA 0.025U #! Test Tol 0.025, Sys Tol 0.006, TUR 4.167 (>= 4.00). 4.001 DISP Set the 1550A MICROSTRAIN buttons to 4,4,4. 4.002 IEEE [@1281]DCI AUTO 4.003 ACC 200 18.208mA 100P% 20P/ 4.004 IEEE [@1281]RDG?[I] 4.005 MEME 4.006 MEMC mA 0.025U #! Test Tol 0.025, Sys Tol 0.0058208, TUR 4.295 (>= 4.00). 5.001 DISP Set the 1550A MICROSTRAIN buttons to 3,3,3. 5.002 IEEE [@1281]DCI AUTO 5.003 ACC 20 14.656mA 100P% 20P/ 5.004 IEEE [@1281]RDG?[I] 5.005 MEME 5.006 MEMC mA 0.025U #! Test Tol 0.025, Sys Tol 0.0018656, TUR 13.401 (>= 4.00). 6.001 DISP Set the 1550A MICROSTRAIN buttons to 2,2,2. 6.002 IEEE [@1281]DCI AUTO 6.003 ACC 20 11.104mA 100P% 20P/ 6.004 IEEE [@1281]RDG?[I] 6.005 MEME 6.006 MEMC mA 0.025U #! Test Tol 0.025, Sys Tol 0.0015104, TUR 16.552 (>= 4.00). 7.001 DISP Set the 1550A MICROSTRAIN buttons to 1,1,1. 7.002 IEEE [@1281]DCI AUTO 7.003 ACC 20 7.552mA 100P% 20P/ 7.004 IEEE [@1281]RDG?[I] 7.005 MEME 7.006 MEMC mA 0.025U #! Test Tol 0.025, Sys Tol 0.0011552, TUR 21.641 (>= 4.00). 8.001 JMP 14.001 8.002 JMP 9.001 8.003 EVAL 9.001 DISP Remove the connections at terminals 11 and 13. Be sure 9.001 DISP to note which wire comes from which terminal. 9.002 DISP ÿÿÿÿÿÿÿÿConnect the UUT as follows: 9.002 DISP terminal 11 ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]Datron Hi 9.002 DISP terminal 12 ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ[16]Datron Lo 9.003 DISP Set the 1550A MICROSTRAIN buttons to 5,0,0. 9.004 IEEE [@1281]DCV AUTO 9.005 ACC 20 10.000V 6P% 0.2P/ 9.006 IEEE [@1281]RDG?[I] 9.007 MEME 9.008 MEMC V 0.005U #! Test Tol 0.025, Sys Tol 0.006, TUR 4.167 (>= 4.00). 10.001 DISP Set the 1550A MICROSTRAIN buttons to 4,4,4. 10.002 ACC 20 8.880V 6P% 0.2P/ 10.003 IEEE [@1281]RDG?[I] 10.004 MEME 10.005 MEMC V 0.005U #! Test Tol 0.025, Sys Tol 0.0058208, TUR 4.295 (>= 4.00). 11.001 DISP Set the 1550A MICROSTRAIN buttons to 3,3,3. 11.002 ACC 20 6.660V 6P% 0.2P/ 11.003 IEEE [@1281]RDG?[I] 11.004 MEME 11.005 MEMC V 0.005U #! Test Tol 0.025, Sys Tol 0.0018656, TUR 13.401 (>= 4.00). 12.001 DISP Set the 1550A MICROSTRAIN buttons to 2,2,2. 12.002 ACC 20 4.440V 6P% 0.2P/ 12.003 IEEE [@1281]RDG?[I] 12.004 MEME 12.005 MEMC V 0.005U #! Test Tol 0.025, Sys Tol 0.0015104, TUR 16.552 (>= 4.00). 13.001 DISP Set the 1550A MICROSTRAIN buttons to 1,1,1. 13.002 ACC 20 2.220V 6P% 0.2P/ 13.003 IEEE [@1281]RDG?[I] 13.004 MEME 13.005 MEMC V 0.005U #! Test Tol 0.025, Sys Tol 0.0011552, TUR 21.641 (>= 4.00). 14.001 DISP Reconnect wires from UUT analoge output. 14.002 DISP Reset all previoulsy noted values to their 14.002 DISP original states. 14.003 JMP 14.005 14.004 DISP Consult manual to reset security code to 14.004 DISP factory default. 14.005 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.